Remote control commands
R&S
®
RTO6
1489User Manual 1801.6687.02 ─ 05
Example: See "Simple frequency and amplitude measurement"
on page
1262.
See
"Creating and reading histograms" on page 1263
Usage: Query only
MEASurement<m>:RESult:STARt? [<MeasType>]
MEASurement<m>:
RESult:STOP? [<MeasType>]
Return the start and stop times of the specified measurement. The parameter defines
the measurement. If no parameter is specified, the result of the main measurement is
returned. The main measurement is defined using MEASurement<m>:MAIN.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Query parameters:
<MeasType> See MEASurement<m>:
RESult:STDDev?
on page 1487.
Usage: Query only
MEASurement<m>:RESult:COUNt?
Returns the number of result groups that are returned by MEASure
ment:RESult:ACTual? HAR.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Query parameters:
<MeasT
ype> HAR
The command is only relevant for harmonic search.
Return values:
Number Number of result groups
Example:
:MEASurement:RESult:ACTual? HAR
99.9;-6.000139236;199.8;-80.701713562;299.7;-15.528377533;399.6;-79.840820312
:MEASurement:
RESult:
COUNt? HAR
4
Usage: Query only
Firmware/software: Version 2.70
24.12.3 Amplitude/time measurements
The following table lists the <MeasType> parameter values with a short description.
Automatic measurements