Remote control commands
R&S
®
RTO6
2565User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:LOAD:REMProbeload <V
alue>
Removes the loading of the probe if DEEMbedding<m>:
COMPonent<n>:MODE is set
to PROBing.
If the probe loading is removed, you see the signal that would be at the measurement
point if the probe's input impedance is ideal (infinite impedance).
Without removing the loading, you see the real signal at the measurement point,
including the probe loading.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1, only for available DUT
Parameters:
<Value> OFF | ON
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:LOAD:IMPedance? <Value>
Shows the source impedance of the DUT
. If an R&S RTZM probe is connected, the
value depends on the selected probe mode: common mode, differential, or single-
ended measurement.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1, only for available DUT
Return values:
<Value> Numeric value
Usage: Query only
Asynchronous command
24.24.3.3 R&S RTZM probe settings
This chapter lists the commands that correspond to the "Deembedding"
> "RT-ZMxx"
dialog box. More commands are provided in Chapter 24.8.7.6, "Modular probes, deem-
bedding", on page 1358.
PROBe<m>:DEEMbedding:TIPModule[:SELect]............................................................. 2566
PROBe<m>:DEEMbedding:TIPModule:ZMA<n>:SUBModule...........................................2566
PROBe<m>:DEEMbedding:BANDwidth..........................................................................2566
Deembedding (option R&S R
T
O6-K121)