Remote control commands
R&S
®
RTO6
2498User Manual 1801.6687.02 ─ 05
Return values:
<CompDirection> UNDefined | VERTical | HOEXtrinsic | HOINtrinsic
HOEXtrinsic: horizontal extrinsic
HOINtrinsic: horizontal intrinsic
*RST: UNDefined
Usage: Query only
ADVJitter<m>:RESult:PERiodic<n>:VALue?
Returns the value of the selected periodic component.
Suffix:
<m>
.
1..2
<n> *
Number of periodic components.
Return values:
<CompValue> Range: -1000 to 1000
Increment: 0
*RST: 0
Usage: Query only
ADVJitter<m>:RESult:PERiodic<n>[:FREQuency]?
Returns the frequency of the selected periodic component.
Suffix:
<m>
.
1..2
<n> *
Number of periodic components.
Return values:
<FreqValue> Range: 0 to 1E+15
Increment: 0
*RST: 0
Default unit: Hz
Usage: Query only
Measurements
ADVJitter<m>:DCOMposition:CLEBasicmeas................................................................. 2499
ADVJitter<m>:DCOMposition:
SELBasicmeas................................................................. 2499
ADVJitter<m>:DCOMposition:CLEJitcomp......................................................................2499
ADVJitter<m>:DCOMposition:SELJitcomp......................................................................2499
ADVJitter<m>:DCOMposition:CLENoisecomp.................................................................2499
ADVJitter<m>:DCOMposition:SELNoisecomp.................................................................2499
ADVJitter<m>:RESult:TRANsitions................................................................................ 2500
ADVJitter<m>:RESult:LEVels........................................................................................ 2500
Jitter analysis and clock data recovery