Triggers
R&S
®
RTO6
220User Manual 1801.6687.02 ─ 05
For information on triggering on serial buses, see the "Trigger" chapter of the relevant
protocol.
6.3.18 T
riggering on parallel buses and digital channels
T
riggering on digital signals requires the Mixed Signal Option. The option is described
in Chapter 14, "Mixed signal option (MSO, R&S RTO6-B1)", on page 984.
For information on triggering, see Chapter 14.3.1, "Trigger settings for digital signals
and parallel buses", on page 997.
6.4 Mode / Holdoff
Access: "Trigger" menu > "Mode / Holdoff" tab
Holdoff conditions define a waiting time after the current trigger until the next trigger
can be recognized.
[Run Stop]/[Single]
Front panel keys to start and stop a continuous acquisition or a defined number of
acquisition cycles, respectively. The number of acquisitions is set with
"Average count".
Remote command:
RUN on page 1325
SINGle on page 1325
STOP on page 1326
Trigger mode
Sets the trigger mode which determines the behavior of the instrument if no trigger
occurs. The current setting is shown on the trigger label on top of the signal bar.
T
o toggle quickly between "Auto" and "Normal" mode, use the [Auto Norm] key on the
front panel (in "Trigger" section).
Mode / Holdoff