Waveform generator (Option R&S RTO6-B6)
R&S
®
RT
O6
1026User Manual 1801.6687.02 ─ 05
Duty cycle
Sets the duty cycle for a square waveform. The duty cycle expresses the percentage of
the period during which the waveform is active, i.e. the signal state is high.
Remote command:
WGENerator<m>:
MODulation:
FM:DCYCle on page 2448
15.1.3.5 FSK modulation
For frequency shift keying (FSK) modulation, the signal switches between Frequency 1
and Frequency 2 at a Hop rate.
Frequency 1
Sets the frequency of the first signal in FSK modulated signal.
Remote command:
WGENerator<m>:MODulation:
FSK:FONE on page 2449
Frequency 2
Sets the frequency of the second signal in FSK modulated signal.
Remote command:
WGENerator<m>:MODulation:FSK:FTWO on page 2450
Hop rate
Sets the hop rate, the frequency at which signal switches between Frequency 1 and
Frequency 2.
Remote command:
WGENerator<m>:MODulation:FSK[:RATE] on page 2450