Remote control commands
R&S
®
RTO6
1617User Manual 1801.6687.02 ─ 05
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
24.15.9 W
idth search conditions
T
rigger level setting: SEARch:TRIGger:LEVel[:VALue] on page 1607
SEARch:TRIGger:WIDTh:DELTa................................................................................... 1617
SEARch:TRIGger:WIDTh:POLarity................................................................................ 1617
SEARch:TRIGger:WIDTh:RANGe..................................................................................1618
SEARch:TRIGger:WIDTh:WIDTh...................................................................................1618
SEARch:TRIGger:WIDTh:DELTa <SearchName>,<WidthDelta>
SEARch:TRIGger:WIDTh:DELTa? <SearchName>
Defines a range around the given width value (see also SEARch:TRIGger:WIDTh:
WIDTh on page 1618).
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 500E-12
*RST: 0
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Example: See Chapter 24.3.4.1, "Searching for a pulse of specified width",
on page
1266
Usage: Asynchronous command
SEARch:TRIGger:WIDTh:POLarity <SearchName>,<Polarity>
SEARch:TRIGger:WIDTh:POLarity? <SearchName>
Indicates the polarity of a pulse, that is the direction of the first pulse slope.
Parameters:
<Polarity> POSitive | NEGative | EITHer
See Chapter 24.4.4, "Polarity parameter", on page
1285.
*RST
: POSitive
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
Search