EasyManua.ls Logo

R&S RTO6 - 14.3.1.6 State

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mixed signal option (MSO, R&S RTO6-B1)
R&S
®
RT
O6
1004User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:
PARallel:
SETHold:HTIMe on page 2436
14.3.1.6 State
The state trigger detects the logical state of several logically combined digital channels
at a given clock edge. The trigger source is a logical combination of digital channels or
a parallel bus. The trigger occurs at the clock edge at which the state condition is true.
Figure 14-5: State trigger settings for trigger source = parallel bus
Clock source
Selects the digital channel of the clock signal.
Remote command:
TRIGger<m>:PARallel:
SETHold:CSOurce[:VALue] on page 2433
TRIGger<m>:PARallel:STATe:CSOurce:VALue on page 2433
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue] on page 2433
Clock edge
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time at which the logical states and the bus value are analyzed.
Remote command:
TRIGger<m>:PARallel:STATe:CSOurce:EDGE on page 2437
Channel states
Available for "Trigger Source"= "Par. Bus1/2/3/4"
For each digital channel that is used in the bus, set the required state: 1, 0, or X (don't
care).
Trigger

Table of Contents

Other manuals for R&S RTO6

Related product manuals