Remote control commands
R&S
®
RTO6
1554User Manual 1801.6687.02 ─ 05
Example:
REFL2:REL:MODE FIVE
Reference levels for C1W1: Lower reference level = 5% of high
signal level, middle reference level = 50% of high signal level,
upper reference level = 95% of high signal level
Example: See: "Manual reference level definition using relative values"
on page 1262
Usage: Asynchronous command
REFLevel<m>:USRLevel
<Mode>
Defines whether the user-defined signal levels or user-defined reference levels are
used for the measurements.
Suffix:
<m>
.
V
alid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Mode> USIGnal | UREF
USIGnal
The high and low signal levels are defined by the user
.
UREF
The reference levels are defined by the user
.
Example:
REFLevel2:USRLevel UREF
Sets user-defined reference levels to be used for C1W1. C1W1
corresponds to suf
fix number 2.
Example: See:
"Manual reference level definition using absolute values"
on page 1262
Usage: Asynchronous command
24.12.15.2 Automatic configuration
REFLevel<m>:AUTO:MODE
......................................................................................... 1554
REFLevel<m>:AUTO[:STATe]........................................................................................1555
REFLevel<m>:AUTO:COUNt........................................................................................ 1556
REFLevel<m>:AUTO:MODE <Mode>
Defines the high and low signal levels from which the reference levels are derived.
This setting is only available for automatic reference level mode (see REFLevel<m>:
LDETection on page 1552).
Automatic measurements