Acquisition and waveform setup
R&S
®
RTO6
180User Manual 1801.6687.02 ─ 05
Max voltage ← External shunt mode
Selects the maximum voltage at the external shunt.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:
MAXVoltage on page 1372
Shunt value ← External shunt mode
Sets the shunt value of the external shunt resistor.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:EVALue on page 1370
Max current ← External shunt mode
Displays the maximum current.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:MXCValue? on page 1371
Reset Overload ← External shunt mode
Resets the overload indication at the probe.
Remote command:
ZVC:Z<m>:I<n>:OVERload:RSTO on page 1369
5.8 Differential signals
With R&S RTO6, you can easily analyze differential signals using single-ended probes,
or even cables. The instrument processes the input of single-ended probes on analog
channels and creates the differential and common mode waveforms. Similar to
Rohde & Schwarz modular probes, you can display the waveforms of differential volt-
age, common mode voltage, positive single-ended voltage and negative single-ended
voltage. Triggering on differential signals is also possible.
Deembedding is possible if option R&S RTO6-K121 is installed.
Deembedding is applied to the two single-ended captured signals, resulting in deem-
bedded differential waveform, common mode waveform, positive single-ended wave-
form and negative single-ended waveform. Deembedding also considers cross-cou-
plings.
Differential signals