Remote control commands
R&S
®
RTO6
2462User Manual 1801.6687.02 ─ 05
24.19.6 Coupling & sync
WGENerator<m>:COUPling:ST
A
Te................................................................................2462
WGENerator<m>:COUPling:AMPLitude......................................................................... 2462
WGENerator<m>:COUPling:PHASeshift........................................................................ 2462
WGENerator<m>:COUPling[:FREQuency]......................................................................2463
WGENerator<m>:COUPling:NOISe............................................................................... 2463
WGENerator<m>:COUPling:AWForm............................................................................ 2463
GENerator:SYNC[:COMBination]...................................................................................2463
WGENerator<m>:COUPling:STATe <CplEnabSt>
Enables the coupling of the generators, with the selected set of parameters: amplitude,
frequency and noise.
Suffix:
<m>
.
1..2
Parameters:
<CplEnabSt> ON | OFF
*RST: OFF
WGENerator<m>:COUPling:AMPLitude <CoupleAmplitude>
Enables the coupling of all amplitude parameters of the generators.
Suffix:
<m>
.
1..2
Specifies the master generator.
Parameters:
<CoupleAmplitude>
ON | OFF
*RST: OFF
WGENerator<m>:COUPling:PHASeshift <PhaseShift>
Sets the phase shift between the waveform of Gen1 and Gen2 when the frequency
parameters of the two waveforms are coupled.
Suffix:
<m>
.
1..2
Specifies the master generator
.
Parameters:
<PhaseShift>
Range: -180 to 180
Increment: 1
*RST: 0
Default unit: °
Waveform generator (option R&S RT
O6-B6)