Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1170User Manual 1801.6687.02 ─ 05
The used resources are listed in the "Details" tab. See also: Chapter 19.2.3, "Details
tab", on page 1135
.
The" Turn on time" is measured as the time between the trigger point ("Trigger level
on" value is reached) and the time the given percentage of the "Steady state level" is
reached.
The" Turn off time" is measured as the time between the trigger point, delayed with the
set "Time" ("Trigger level on" value is reached) and the time the given percentage of
the "Steady state level" is reached, see
See also Figure 19-2 and Figure 19-3
The following remote commands are used for handling the measurement results:
POWer:ONOFf:RESult:TOFF? on page 2543
POWer:ONOFf:RESult:TON? on page 2543
POWer:ONOFf:REPort:ADD on page 2542
19.5.2.2 Configuring turn on/ off
For details of the configuration settings, see Chapter 19.5.2.3, "Turn on/off settings",
on page 1172.
1. Select "Analysis">"Power".
2. Under "Power Path", select "Turn on/off ".
3. Connect the probes to the DUT as shown in the "Channels" tab:
Power path