Remote control commands
R&S
®
RTO6
2564User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:PCOunt?
Returns the number of all ports of the selected component (input ports and output
ports).
The number of ports depends on the attached probe. If a modular probe R&S R
T
‑ZM
or differential probe R&S RT-ZD is used, the number of ports is 4. Otherwise, 2 ports
are available.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..n
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Return values:
<TotalPorts> ONE | TWO | THRee | FOUR
*RST: TWO
Usage: Query only
Asynchronous command
24.24.3.2 DUT settings
DEEMbedding<m>:COMPonent<n>:MODE
....................................................................2564
DEEMbedding<m>:COMPonent<n>:LOAD:REMProbeload..............................................2565
DEEMbedding<m>:COMPonent<n>:LOAD:IMPedance?..................................................2565
DEEMbedding<m>:COMPonent<n>:MODE <Mode>
Sets the probing mode.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1, only for available DUT
Parameters:
<Mode> PROBing | TERMinating
PROBing
For measurements with a high-ohmic probe and measure on an
existing line in parallel to the load.
TERMinating
For measurements if the measuring equipment is the load of the
line.
*RST: PROBing
Deembedding (option R&S RT
O6-K121)