Remote control commands
R&S
®
RTO6
2563User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 2..n-1
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Parameters:
<IsEnabled> ON | OFF
*RST: OFF
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:TYPE <Type>
Sets the type of the indicated component. The first component is always DUT, the last
component is PROBe or INPut.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..n
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Parameters:
<T
ype> PROBe | FIXTure | CABLe | DUT | ADAPter | INPut | OTHer |
R
TZA10 | R
TZA16 | RTZA17 | PP | PC
RTZA10 | RTZA16 | RTZA17
SMA adapters and cable offered by Rohde & Schwarz
PP | PC
Proven probe, proven cable
*RST: CABLe
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:NAME <Name>
Sets a name for the indicated component that helps you identifying it.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..n
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Parameters:
<Name> String parameter, name of the component
Deembedding (option R&S RT
O6-K121)