Protocol analysis
R&S
®
RTO6
831User Manual 1801.6687.02 ─ 05
TRIGger<m>:USB:PMIN on page 2158
TRIGger<m>:
USB:PMAX
on page 2158
SEU
Defines, whether an SEU check that meets specific conditions is executed or not. S
and E represent the Start and End of a start-split transaction, U represents the
reserved/Unused bit of a complete-split transaction. Permissible binary SEU values are
00, 01, 10, and 11.
"Condition"
Defining a specific SEU value or an SEU value range requires to set
the operator to one of the following: equal, not equal, less than, less
than or equal, greater than, greater than or equal, in range, out of
range.
"Min"
Specifies an SEU value, or sets the start value of an SEU value
range.
"Max"
Sets the end value of an SEU value range if "Condition"
(TRIGger<m>:
USB:SCONdition) is set to
INRange or OORange.
Remote command:
TRIGger<m>:USB:WSEU
on page 2162
TRIGger<m>:USB:SCONdition on page 2158
TRIGger<m>:USB:SMIN on page 2158
TRIGger<m>:USB:SMAX on page 2159
ET
Defines, whether an Endpoint Type (ET) check that meets specific conditions is execu-
ted or not.
"Condition"
Defining a specific endpoint type or an ET range requires to set the
operator to one of the following: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies an ET, or sets the start value of an ET range.
"Max"
Sets the end value of an ET range if "Condition" (TRIGger<m>:USB:
TCONdition) is set to INRange or
OORange.
Remote command:
TRIGger<m>:USB:WETCheck
on page 2161
TRIGger<m>:USB:TCONdition on page 2160
TRIGger<m>:USB:TMIN on page 2160
TRIGger<m>:USB:TMAX on page 2160
13.20.3.2 Triggering on USB 2.0
Prerequisite: A bus is configured for the USB 2.0 signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1.
Press the [Bus]
key and selet the "Shortcuts" tab.
2. Press "Setup trigger".
USB (option R&S RT
O6-K570)