Protocol analysis
R&S
®
RTO6
745User Manual 1801.6687.02 ─ 05
●
Vil(max) is being used for the falling edge evaluation. This "l" (low) threshold is the
maximum level for the signal to be identified as "0". If the signal value comes from
a high state (hence, falling edge), the state remains to be considered as high ("1"),
until it has fallen below V
il(max).
There are four ways to set the thresholds for the digitization of the signal lines:
●
"Threshold"
Enter the values directly in the fields.
●
"Preset thresholds"
Allows to select the default threshold settings according to the Ethernet standard:
2.0 V and 0.8 V.
●
"Auto thresholds"
Executes a measurement of reference levels and sets the thresholds to the middle
reference level of the measured amplitude.
●
"Couple thresholds"
Overwrites the data thresholds with the clock thresholds.
Remote command:
BUS<m>:MDIO:CLOCk:THReshold:HIGH on page 2041
BUS<m>:MDIO:CLOCk:THReshold:LOW on page 2041
BUS<m>:MDIO:DATA:THReshold:HIGH on page 2041
BUS<m>:MDIO:DATA:THReshold:LOW on page 2042
BUS<m>:MDIO:PRESet on page 2042
BUS<m>:SETReflevels on page 1671
BUS<m>:MDIO:COUPling on page 2042
13.16.2.2 MDIO display settings
Access: [Bus] > "Setup" tab > "Protocol = MDIO" > "Display" tab
T
o enhance the decode possibilities of the MDIO protocol, you can use an additional
setting in the "Display" tab: "Decode layer".
MDIO (option R&S RT
O6-K540)