Jitter analysis and clock data recovery
R&S
®
RTO6
1074User Manual 1801.6687.02 ─ 05
... ... ...
N periods N periods
T
Ncycle1
T
Ncycle2
ΔT
Ncycle1
= T
Ncycle2
- T
Ncycle1
Remote command:
MEASurement<m>:JITTer:NCYCles on page 2481
18.1.2.3 Data measurement settings
The following measurements are intended to analyze serial data. The clock can be a
captured clock signal, or it can be recovered from the data signal using one of the
clock recovery algorithms:
Time interval error
Unit interval
Data rate
To configure a measurement, tap it in the "Active Measurements" list.
Figure 18-1: Settings for time interval error measurement with real clock signal
Figure 18-2: Settings for data rate measurement with software clock recovery
Jitter measurements (options R&S RT
O6-K12/K13)