Acquisition and waveform setup
R&S
®
RTO6
164User Manual 1801.6687.02 ─ 05
Offset to mean
Performs an automatic compensation for a DC component of the input signal using the
result of a background mean measurement. The result is shown in
"Of
fset". The func-
tion is probe-independent and supports quick and convenient measurements of input
signals with different DC offsets. It detects offset values even when the signal is out of
the current measurement range. It also sets the zero level to the determined DC offset
in the middle of the screen and thus prevents clipping of the waveform.
Remote command:
PROBe<m>:SETup:OFFSet:TOMean on page 1362
Line impedance
If the transmission line probe R&S RT-ZZ80 is selected, enter the impedance of the
measured line.
The actual attenuation of the transmission line probe depends on the impedance of the
line Z
0
:
Attenuation = 10 + Z
0
/ 100
The instrument uses the actual attenuation to determine the measurement values.
5.6.7 Setup for current probes
The setup and adjustment of current probes depends on the output connector of the
probe: BNC or Rohde & Schwarz probe box.
The following shared probe settings are available:
"Type, Name, Bandwidth" on page 146
"Detect AutoZero, Use AutoZero" on page 147
Current probes R&S RT-ZCxx
The current probes R&S RT-ZCxx have BNC connectors. They are known to the
R&S RTO6 as predefined probes, see Chapter 5.6.6, "Setup for predefined probes",
on page 162. Demagnetizing and zero adjustment is done on the probe, see the pro-
be's User Manual for details. Make sure to demagnetize and adjust the probe before
taking measurements.