Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1196User Manual 1801.6687.02 ─ 05
The following remote commands are used for handling the measurement results:
●
POWer:SPECtrum:
RCOunt?
on page 2558
●
POWer:SPECtrum:RESult<m>:FREQuency? on page 2558
●
POWer:SPECtrum:RESult<m>:LEVel? on page 2558
●
POWer:SPECtrum:REPort:ADD on page 2558
19.6.3.2 Configuring output spectrum
For details of the configuration settings, see Chapter 19.6.3.3, "Output spectrum set-
tings", on page 1197.
1. Select "Analysis">"Power".
2. Under "Output", select "Spectrum".
3. Select the "Channels" tab.
4. Connect the probe to the DUT and to the oscilloscope as shown in the graphic:
5. Select the correct channel for the "Voltage Source".
6. Select the "Settings" tab.
7. Set the "SMPS switching frequency" according to your signal.
8. Select an "Optimize Scaling"option.
Output