Remote control commands
R&S
®
RTO6
1398User Manual 1801.6687.02 ─ 05
Parameters:
<WidthDelta> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-12
Default unit: s
Usage: Asynchronous command
24.9.6 Window trigger
TRIGger<m>:LEV
el<n>:WINDow:LOWer........................................................................1398
TRIGger<m>:LEVel<n>:WINDow:UPPer........................................................................ 1398
TRIGger<m>:WINDow:RANGe......................................................................................1398
TRIGger<m>:WINDow:TIME.........................................................................................1399
TRIGger<m>:WINDow:WIDTh.......................................................................................1400
TRIGger<m>:WINDow:DELTa....................................................................................... 1400
TRIGger<m>:LEVel<n>:WINDow:LOWer <Level>
TRIGger<m>:LEVel<n>:WINDow:UPPer <Level>
Set the lower and upper voltage limits for the window.
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
<n> 1..31
Indicates the trigger source:
1...4 = channel 1...4
5...27 = not available
28...31 = differential signals Dif
f1, Diff2 and Common1, Com-
mon2
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: Lower = -0.1, upper =0.1
Default unit: V
Usage: Asynchronous command
TRIGger<m>:WINDow:RANGe <RangeMode>
Defines the signal run in relation to the window:
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<RangeMode> ENTer | EXIT | WITHin | OUTSide
Trigger