Remote control commands
R&S
®
RTO6
1397User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Mode> ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
T
riggers on runts longer than the given "Runt width".
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "±Delta".
OUTSide
Triggers if the runt length is outside a given time range. The
range is defined by "Runt width" and "±Delta".
*RST: ANY
Usage: Asynchronous command
TRIGger<m>:RUNT:WIDTh <Width>
Defines the upper or lower voltage threshold. This command is not available if
TRIGger<m>:RUNT:
RANGe is set to "ANY".
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:RUNT:DELTa <WidthDelta>
Defines a range around the runt width specified using TRIGger<m>:RUNT:WIDTh.
This command is only available if TRIGger<m>:
RUNT:RANGe is set to "WITHin" or
"OUTSide".
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Trigger