Acquisition and waveform setup
R&S
®
RTO6
141User Manual 1801.6687.02 ─ 05
You can use inversion, for example, to switch the polarity of a differential signal without
changing the probe connections.
Remote command:
CHANnel<m>:
INVert on page
1339
Ground
Connects the input to the ground.
Remote command:
CHANnel<m>:GND on page 1338
5.5.2 Bandwidth settings
Access: "Menu" > "Vertical" > "Bandwidth"
The "Bandwidth" tab provides all settings that affect the bandwidth of the measurement
system. The channels are listed in horizontal subtabs. Make sure to select the correct
channel tab before you enter the settings.
Analog
Selects the bandwidth limit.
The specified bandwidth indicates the range of frequencies that the instrument can
acquire and display accurately with less than 3dB attenuation. The probe has also a
limited bandwidth and thus af
fects the resulting system bandwidth.
See also:
Chapter 5.1.1.3, "Bandwidth", on page 111
"Full"
At full bandwidth, all frequencies in the specified range are acquired
and displayed. Full bandwidth is used for most applications.
Vertical setup