Protocol analysis
R&S
®
RTO6
721User Manual 1801.6687.02 ─ 05
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:SWIRe:
DATA:CONDition on page 2003
TRIGger<m>:SWIRe:DATA:MAX on page 2003
TRIGger<m>:SWIRe:DATA:MIN on page 2004
Parity Error
Checks the parity of every frame and triggers if the parity is even.
Remote command:
TRIGger<m>:SWIRe:ERRor:PARity on page 2004
ESC Error
Triggers on an escape error.
Remote command:
TRIGger<m>:SWIRe:ERRor:ESC on page 2004
13.14.3.2 Triggering on SpaceWire
Prerequisite: A bus is configured for the SpaceWire signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [Bus] key and select the "Shortcuts" tab.
2. Press "Setup trigger".
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to SpaceWire.
5. Select the "Trigger Type" to be used for SpaceWire protocol analysis.
6. To refine the trigger settings, configure additional settings, which are available for
some trigger types.
For details, see Chapter 13.14.3.1, "SpaceWire trigger settings", on page 719.
13.14.4 SpaceWire decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. In the "Protocol" dialog > "Setup" tab, enable "Decode".
2. In the "Protocol" dialog > "Display" tab, select additional result display settings:
"Show decode table" and "Show binary signals".
For a description of the display settings, see also Chapter 13.1.3, "Display",
on page 487.
SpaceWire (option R&S R
T