Remote control commands
R&S
®
RTO6
2393User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
<n> *
Return values:
<State> OK | UNKNown | INComplete | LENGth
*RST: OK
Usage: Query only
BUS<m>:DDR:RESult:FRAMe<n>:STOP?
Returns the stop time of the frame with the selected frame number (for the selected
serial bus).
Suffix:
<m>
.
1..4
<n> *
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:DDR:RESult:FRAMe<n>:TYPE?
Returns the frame type for the selected serial bus and frame number.
Suffix:
<m>
.
1..4
<n> *
Usage: Query only
24.17.28.3 Search settings
There are two commands for each parameter
, that you can use for defining the DDR
search settings.
For example, to set the
Frame type =Read > Field =Data >Data value you can use one
of the following commands:
SEARch:TRIGger:DDR:FRAMe1:ENABle:DMIN 01100
Defines the parameter by using the index <m> for the frame number and <n> for
the field number.
SEARch:TRIGger:DDR:DMIN "Read", "Data", 01100
Defines the parameter by using the frame and field name.