Protocol analysis
R&S
®
RTO6
616User Manual 1801.6687.02 ─ 05
Tools
Press "Display settings"
in the "Decode Results"
table to access the "Display" tab.
Remote commands
Remote commands are described in Chapter 24.17.9.3, "Decode results",
on page
1845.
13.8.5
Search on decoded LIN data
Using the search functionality, you can find various events in the decoded data, the
same events which you also can trigger on. Before you can start the search, you have
to configure the bus correctly and acquire decoded data.
To search on decoded data, set the search "Source" to the serial bus that is configured
for the protocol to be analyzed.
See also Chapter 11, "Search functions", on page 419.
13.8.5.1 LIN search setup
Access: [Bus] > "Shortcuts" tab > "Setup search"
Search types
Sets the type to be searched for.
"Start of frame
(Sync)"
Searches for the stop bit of the sync field.
"Identifier"
Searches for one specific identifier or an identifier range.
See
"Identifier setup: Condition, Frame ID min, Frame ID max"
on page 617
"Identifier OR"
Searches for a combination of up to four identifiers.
See
"Identifier OR setup: Monitor, Frame ID"
on page 617
LIN (option R&S RT
O6-K520)