Protocol analysis
R&S
®
RTO6
798User Manual 1801.6687.02 ─ 05
13.19.2.3 Configuring 1000BASE-T1 signals
For configuration, assign the lines to the input channels and define the active states
and the logical thresholds.
Serial bus setup
1.
Press the [Bus]
key on the front panel.
2. Select the tab of the bus you want to set up, for example "SB1".
3. In the "Setup" tab, select the protocol: "1000BASE-T1".
4. Select the type and source for the reversed and forward signal.
5. Select the "Polarity" ("Normal" or "Inverted") of the signal.
6. Enable "Decode".
13.19.3 1000BASE-T1 trigger
13.19.3.1 1000BASE-T1 trigger settings
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
Ethernet 1000BASE-T1 (option R&S RT
O6-K560)