Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1187User Manual 1801.6687.02 ─ 05
POWer:RIPPle:RESult:
PERiod:EVTCount? on page 2553
POWer:RIPPle:RESult:PERiod:NPEak? on page 2553
POWer:RIPPle:RESult:PERiod:PPEak? on page 2553
POWer:RIPPle:RESult:PERiod:RMS? on page 2553
POWer:RIPPle:RESult:PERiod:STDDev? on page 2553
POWer:RIPPle:RESult:PERiod:WFMCount? on page 2553
POWer:RIPPle:RESult:PERiod[:ACTual]? on page 2553
POWer:RIPPle:RESult:STDDev:AVG? on page 2554
POWer:RIPPle:RESult:STDDev:EVTCount? on page 2554
POWer:RIPPle:RESult:STDDev:NPEak? on page 2554
POWer:RIPPle:RESult:STDDev:PPEak? on page 2554
POWer:RIPPle:RESult:STDDev:RMS? on page 2554
POWer:RIPPle:RESult:STDDev:STDDev? on page 2554
POWer:RIPPle:RESult:STDDev:WFMCount? on page 2554
POWer:RIPPle:RESult:STDDev[:ACTual]? on page 2554
POWer:RIPPle:REPort:ADD on page 2550
19.6.1.2 Configuring output ripple
For details of the configuration settings, see Chapter 19.6.1.3, "Output ripple settings",
on page 1188.
1. Select "Analysis" > "Power".
2. Under "Output", select "Ripple".
3. Connect the voltage probe to the oscilloscope.
4. If you want to measure the current ripple, enable the current channel .Connect the
current probe to the oscilloscope.
5. If you want to measure both the voltage and the current ripple, deskew the probes
as described in Chapter 19.1.1.1, "Auto deskew", on page 1123.
6. Connect the probes to the DUT as shown in the "Channels" tab:
Output