Remote control commands
R&S
®
RTO6
1394User Manual 1801.6687.02 ─ 05
TRIGger<m>:GLITch:WIDTh <Width>
Sets the length of a glitch. The instrument triggers on pulses shorter or longer than this
value, depnding on the TRIGger<m>:
GLITch:RANGe command.
Y
ou need to know the expected pulse widths of the circuit to set the glitch width cor-
rectly.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-6
*RST: 1E-9
Default unit: s
Usage: Asynchronous command
24.9.4 Width trigger
TRIGger<m>:WIDTh:
POLarity.......................................................................................1394
TRIGger<m>:WIDTh:RANGe........................................................................................ 1394
TRIGger<m>:WIDTh:WIDTh......................................................................................... 1395
TRIGger<m>:WIDTh:DELTa..........................................................................................1395
TRIGger<m>:WIDTh:POLarity <Polarity>
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Polarity> POSitive | NEGative
See Chapter 24.4.4, "Polarity parameter", on page 1285.
*RST
: POSitive
Usage: Asynchronous command
TRIGger<m>:WIDTh:RANGe <RangeMode>
Defines how the range of a pulse width is defined in relation to the width and delta
specified using TRIGger<m>:WIDTh:
WIDTh and TRIGger<m>:WIDTh:DELTa,
respectively.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
Trigger