EasyManua.ls Logo

R&S RTO6 - 13.11.3.2 Triggering on CXPI

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
678User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:
CXPI:
DATA:DCONdition on page 1934
TRIGger<m>:CXPI:DATA:DMAX on page 1934
TRIGger<m>:CXPI:DATA:DMIN on page 1935
Index
The data index setup consists of the condition and one or two index patterns.
"Condition"
Defines the operator to set a specific data index ("Equal" or "Not
equal") or data index range.
"Index min /
Index"
Defines the bit pattern of the data index.
"Index max"
The second data index is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
TRIGger<m>:CXPI:
DATA:ICONdition on page 1935
TRIGger<m>:CXPI:DATA:IMAX on page 1935
TRIGger<m>:CXPI:DATA:IMIN on page 1935
NM
Sets the value of the network management (NM) field, 2 bits indicating if the frame is in
wakeup or sleep mode.
Remote command:
SEARch:TRIGger:CXPI:NM on page 1952
CT
Sets the value of the counter (CT), 2 bits indicating the continuity of the frame.
Remote command:
TRIGger<m>:CXPI:CT on page 1934
13.11.3.2 Triggering on CXPI
Prerequisite: A bus is configured for the CXPI signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [Bus]
key and select the "Shortcuts" tab.
2. Press "Setup trigger".
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to CXPI.
5. Tap "Trigger Type" and select the trigger type to be used for CXPI protocol analy-
sis.
6. Depending on the selected trigger type, more setup conditions can be specified.
CXPI (option R&S RTO6-K520)

Table of Contents

Other manuals for R&S RTO6

Related product manuals