R&S
®
RT
O6
1184User Manual 1801.6687.02 ─ 05
The measurement also includes the AC-RMS of the output DC signal, that is calculated
as a standard derivation.
Required probes:
V
oltage probe
(Optional) Current probe
19.6.1.1
Output ripple results
The results of "Output Ripple" measurements are provided in two ways:
The diagram shows the graphical presentation of:
the voltage waveform
(optional) the current waveform
The result box displays the numeric measurement results for the voltage and for
the current ripple. For details, see Table 19-5 and Table 19-2.
To measure and display the output ripple, the instrument uses the following measure-
ments and waveforms:
"MG 7" to measure the current
"MG 8"
to measure the voltage
The used resources are listed in the "Details" tab. See also: Chapter 19.2.3, "Details
tab", on page 1135.
Table 19-5: Properties output ripple
Mea
s.
type
Symbol Description/Result
Max X
Max
Absolute maximum value of the waveform.
Min X
Min
Absolute minimum value of the waveform.
Output