Remote control commands
R&S
®
RTO6
2457User Manual 1801.6687.02 ─ 05
Parameters:
<FilePath> String with the name of the file.
PGENerator:HLEVel <HighLevel>
Sets the high level of the pattern generator signal.
Parameters:
<HighLevel> Range: 1.2 to 5
Increment: 0.1
*RST: 1.2
Default unit: V
PGENerator:PRESet
Sets the default pattern generator settings.
Usage: Event
PGENerator:RUNMode <RunMode>
Sets the duration for which the signal of the pattern generator will be output after the
trigger event.
Parameters:
<RunMode> SINGle | CONTinuous
*RST: CONTinuous
PGENerator:RUNSingle
Executes a single period of the pattern generator, for PGENerator:
RUNModeSINGle.
Usage: Event
24.19.3 Output setup
WGENerator<m>:OUTPut[:
LOAD]
................................................................................. 2457
WGENerator<m>:VOLTage[:VPP]..................................................................................2458
WGENerator<m>:VOLTage:DCLevel..............................................................................2458
WGENerator<m>:VOLTage:HIGH.................................................................................. 2458
WGENerator<m>:VOLTage:INVersion............................................................................ 2459
WGENerator<m>:VOLTage:LOW...................................................................................2459
WGENerator<m>:VOLTage:OFFSet...............................................................................2459
WGENerator<m>:OUTPut[:LOAD] <Load>
Sets the load of the DUT at its connection.
Waveform generator (option R&S RT
O6-B6)