WGENerator<m>:FUNCtion:PULSe[:WIDTh] <PulseWidth>
Sets the pulse duration for a pulse waveform.
Suffix:
<m>
.
1..2
Parameters:
<PulseWidth> Range: 16.5E-9 to 1000, actual range depends on the
selected frequency
Increment: 1
*RST: 500E-9
Default unit: s
WGENerator<m>:FUNCtion:RAMP[:SYMMetry] <RampSymmetry>
Sets the symmetry for a ramp waveform.
Suffix:
<m>
.
1..2
Parameters:
<RampSymmetry> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
WGENerator<m>:FUNCtion:SQUare:DCYCle <SquareDutyCycle>
Sets the duty cycle for the generated square waveform.
Suffix:
<m>
.
1..2
Parameters:
<SquareDutyCycle> Range: 0.01 to 99.99 at minimum frequency, actual range
depends on the selected frequency
Increment: 1
*RST: 50
Default unit: %
24.19.1.3 Modulation
WGENerator<m>:MODulation:TYPE
..............................................................................2446
WGENerator<m>:MODulation:AM:DCYCle.....................................................................2446
WGENerator<m>:MODulation:AM:DEPTh...................................................................... 2446
WGENerator<m>:MODulation:AM:FREQuency...............................................................2447
WGENerator<m>:MODulation:AM:SYMMetry................................................................. 2447
WGENerator<m>:MODulation:CARRier:FREQuency.......................................................2447
WGENerator<m>:MODulation:CARRier:PERiod..............................................................2447
WGENerator<m>:MODulation:AM[:FUNCtion].................................................................2448