R&S RTO6 - 8.2.13 Limit and margin checks

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements
R&S
®
RTO6
363User Manual 1801.6687.02 ─ 05
Remote command:
MEASurement<m>:
STATistics:
MODE on page 1535
Time / point
Defines the time to create one point of the long-term measurement. The "Time / point"
value is a lower time limit. The actual time between two points depends on the acquisi-
tion and postprocessing time.
The long-term measurement is not a data logger with equidistant points as the time
between two points varies.
This setting is only available if "Scale mode" is set to "Time".
Remote command:
MEASurement<m>:STATistics:RTIMe on page 1536
Measurement time
Defines the total duration of the long-term measurement: Time/point * Number of
points.
This setting is only available if "Scale mode" is set to "Time".
Remote command:
MEASurement<m>:LTMeas:TIME on page 1539
Wfms / point
Defines the number of measured waveforms from which one point of the long-term
measurement is created.
This setting is only available if "Scale mode" is set to "Waveforms".
Remote command:
MEASurement<m>:STATistics:RCOunt on page 1535
Measurements / point
Defines the number of measurement results from which one point of the long-term
measurement is created.
This setting is only available if "Scale mode" is set to "Measurements".
Remote command:
MEASurement<m>:STATistics:RMEascount on page 1536
8.2.13 Limit and margin checks
Limit and margin checks evaluate whether the measurement result exceeds a specified
value. Violations are indicated by icons in the result box. Furthermore, you can define
actions that are performed on limit or margin violation, like saving the waveform or
measurement results.
The following results are considered in limit and margin checks:
All selected measurements.
All measured events in an acquisition: all results can initiate an action. However,
the icons in the result box indicate only violations of the first result.
Automatic measurements