EasyManua.ls Logo

R&S RTO6 - 5.2 Horizontal; 5.2.1 Setup settings

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquisition and waveform setup
R&S
®
RTO6
119User Manual 1801.6687.02 ─ 05
Voltage between the positive and negative signal sockets
Positive single-ended input voltage (V
p
)
Voltage between the positive signal socket and the ground socket
Negative single-ended input voltage (V
n
)
V
oltage between the negative signal socket and the ground socket
Common mode input voltage (V
cm
)
Mean voltage of positive and negative signal sockets referred to the ground socket,
respectively
Two of these voltages are independent values, the other two can be calculated:
2
np
cm
npin
VV
V
VVV
R&S RT-ZD probes detect only differential input voltages and provide it to the oscillo-
scope. Common mode signals are suppressed by the probe. This characteristic is
described by the common mode rejection ratio (CMRR):
GainCommonMode
alGainDifferenti
CMRR
In addition, the R&S ProbeMeter of R&S RT-ZD differential probes can measure differ-
ential and common mode DC voltages. The measurement result is displayed on the
oscilloscope's screen. The common mode measurement of the R&S ProbeMeter
allows you to check the input voltage relative to ground. Thus, the CM measurement is
a convenient way to detect breaches of the operating voltage window
, and the reason
of unwanted clippings.
5.2
Horizontal
The "Horizontal" dialog provides the time base configuration for channel and spectrum
waveforms.
5.2.1 Setup settings
Access: "Menu" > "Horizontal" > "Setup" tab.
Horizontal

Table of Contents