24.17.1 Configuration settings for all serial protocols
BUS<m>:TYPE............................................................................................................
1670
BUS<m>[:ST
ATe]......................................................................................................... 1670
BUS<m>:SETReflevels.................................................................................................1671
BUS<m>:FAUToset...................................................................................................... 1671
BUS<m>:RESult.......................................................................................................... 1671
BUS<m>:THReshold....................................................................................................1671
BUS<m>:FORMat........................................................................................................1672
BUS<m>:NEWList........................................................................................................1672
BUS<m>:SYMBols.......................................................................................................1672
BUS<m>:ZCOupling.....................................................................................................1672
DISPlay:RESultboxes:DEPosition..................................................................................1673
BUS<m>:EXPResult:DETail.......................................................................................... 1673
BUS<m>:EXPResult:SAVE........................................................................................... 1673
BUS<m>:EXPResult:TIME............................................................................................1673
BUS<m>:TYPE <Type>
Defines the bus or protocol type for analysis. The type of available buses depends on
the installed options.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Type> I2C | SPI | UART | CAN | CANFd | LIN | MILS1553 |
MILStd1553 | ARIN429 | ARINc429 | SWIRe | MDIO |
HBT
O |
USB | USBPD | ETHernet | CMSB | FLXRay | I2S | SENT | CXPI
| RFFE | DPHY | MPHY | EBTB | USB3 | PCIE | TBTO
MILS1553 = MILStd1553: specification MIL-STD-1553
ARIN429 = ARINc429: specification ARINC 429
HBTO: Ethernet 100BASE-T1
CMSB: custom decode serial bus
SWIRe: SpaceWire
*RST: I2C
Usage: Asynchronous command
BUS<m>[:STATe] <State>
Enables the decoding of the specified bus.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<State> ON | OFF
*RST: OFF