EasyManua.ls Logo

R&S RTO6 - 15.1.3.4 FM modulation

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Waveform generator (Option R&S RTO6-B6)
R&S
®
RT
O6
1024User Manual 1801.6687.02 ─ 05
Remote command:
WGENerator<m>:
MODulation:
PWM[:FUNCtion] on page 2451
Modulation freq
Sets the frequency of the modulation waveform.
Remote command:
WGENerator<m>:MODulation:PWM:FREQuency on page 2451
Modulation depth
Sets the modulation depth, the percentage of the pulse width range that is used for the
modulation.
Remote command:
WGENerator<m>:MODulation:PWM:DEPTh on page 2450
Symmetry
Sets the symmetry for the ramp modulation waveform, the percentage of time that the
waveform is rising.
Remote command:
WGENerator<m>:MODulation:PWM:SYMMetry on page 2451
Duty cycle
Sets the duty cycle for a square waveform. The duty cycle expresses for what percent-
age fraction of the period, the waveform is active, i.e. the signal state is high.
Remote command:
WGENerator<m>:MODulation:PWM:DCYCle on page 2450
15.1.3.4 FM modulation
For frequency modulation (FM), the frequency of the carrier signal is varied according
to the modulation signal.
Setup of the waveform generator