EasyManua.ls Logo

R&S RTO6 - 24.3.8 Power analysis (option R&S RTO6-K31); 24.3.8.1 Auto deskew

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1277User Manual 1801.6687.02 ─ 05
SEARch:RESult:SENT:FCOunt? 'Search1'
SEARch:RESult:SENT:FRAMe1:STAT? 'Search1'
SEARch:RESult:SENT:FRAMe1:IDValue? 'Search1'
SEARch:RESult:SENT:FRAMe1:NIBB1:VALue? 'Search1'
SEARch:RESult:SENT:FRAMe1:NIBB2:VALue? 'Search1'
SEARch:RESult:SENT:FRAMe1:CSValue? 'Search1'
SEARch:TRIGger:SENT:ENHanced 'Search1', OFF
24.3.8 Power analysis (option R&S RTO6-K31)
24.3.8.1
Auto deskew
Configures the voltage and current probes for power measurements and executes the
auto deskew.
Command description in Chapter 24.23.1, "General", on page 2520 and Chap-
ter 24.23.2, "Deskew", on page 2520.
If the instrument refuses to accept POWer commands, activate the power mode
using :POWer:ENABle.
In the following example *OPC prevents overlapping execution of asynchronous com-
mands.
*RST; *OPC?
STOP;*OPC?
//Activate two channels
:CHANnel1:STATe 1
:CHANnel2:STATe 1
//Activate power
:POWer:ENABle
//Select current and voltage sources
:POWer:SOURce:VOLTage1 CHANnel1
:POWer:SOURce:CURRent1 CHANnel2
*OPC?
//Configure voltage probe on CH1 manually
//Not necessary if you use an active R&S voltage probe that is recognized by the instrument
:PROBe1:SETup:ATTenuation:MODE Manual
:PROBe:SETup:ATTenuation:DEFProbe ZD01a100
*OPC?
//selected high voltage differential probe 1:100
Programming examples