:POWer:ENABle
//Expected smps frequency
:POW:TRANsient:FREQ 12500000
:POWer:TRANsient:AUToscale AUTO
:POWer:TRANsient:SIGHigh 0.1
:POWer:TRANsient:SIGLow 0.025
:POWer:TRANsient:HYSTeresis 20
//Trigger channel
:POWer:TRANsient:TRGC CHAN2
//Edge trigger slope
:POWer:TRANsient:TRGS POS
//Trigger level
:POWer:TRANsient:TRGL 0.08
//Run measurement
:POWer:TRANsient:EXECute;*OPC?
//Query results
:POWer:TRANsient:RESult? SETTling
:POWer:TRANsient:RESult? PEAKtime
:POWer:TRANsient:RESult? DELay
:POWer:TRANsient:RESult? RTIMe
:POWer:TRANsient:RESult? OVERshoot
//Add to report
:Power:TRANsient:REPort:Add
24.4 Frequently used parameters and suffixes
This chapter describes in general those parameters and suf
fixes that are used in sev-
eral subsystems.
24.4.1
Waveform suffix
The numeric waveform suffix is used in some commands, for example, to indicate the
source waveform number from which the reference level is taken, and to assign color
tables to waveforms.
Frequently used parameters and suffixes