HIGH
The signal level stays above the trigger level.
LOW
The signal level stays below the trigger level.
EITHer
The signal level stays above or below the trigger level.
*RST: HIGH
Usage: Asynchronous command
TRIGger<m>:TIMeout:TIME <Time>
Defines the time limit for the timeout at which the instrument triggers.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
24.9.8 Interval trigger
TRIGger<m>:INTerval:SLOPe
....................................................................................... 1401
TRIGger<m>:INTerval:RANGe.......................................................................................1402
TRIGger<m>:INTerval:WIDTh........................................................................................1402
TRIGger<m>:INTerval:DELTa........................................................................................ 1402
TRIGger<m>:INTerval:SLOPe <Slope>
Sets the edge for the trigger. You can analyze the interval between positive edges or
between negative edges.
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative
See Chapter 24.4.3, "Slope parameter", on page 1285.
*RST
: POSitive
Usage: Asynchronous command