Remote control commands
R&S
®
RTO6
1402User Manual 1801.6687.02 ─ 05
TRIGger<m>:INTerval:RANGe <RangeMode>
Defines the range of an interval in relation to the interval width specified using
TRIGger<m>:
INTerval:
WIDTh and TRIGger<m>:INTerval:DELTa.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers on pulses inside a given range. The range is defined by
the interval width ±
delta.
OUTSide
Triggers on pulses outside a given range. The range is defined
by the interval width ±delta.
SHORter
Triggers on pulses shorter than the given interval width.
LONGer
Triggers on pulses longer than the given interval width.
*RST: OUTSide
Usage: Asynchronous command
TRIGger<m>:INTerval:WIDTh <Width>
Defines the time between two pulses.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:INTerval:DELTa <WidthDelta>
Defines a range around the "Interval width"
value specified using TRIGger<m>:
INTerval:WIDTh on page 1402.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Trigger