Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1138User Manual 1801.6687.02 ─ 05
19.3.1.1 Power quality results
The results of " Power Quality"
measurements are provided in two ways:
The diagram shows the graphical presentation of:
the voltage waveform
the current waveform
the power waveform that is the product of the current and voltage waveforms
The result box displays the numeric measurement results.
To measure and display the power quality, the instrument uses the following measure-
ments and waveforms:
"MG 6" to measure the voltage
"MG 7"
to measure the current
"MG 8" to measure the power
"M2" "Math 2" to calculate the power
The used resources are listed in the "Details" tab. See also: Chapter 19.2.3, "Details
tab", on page 1135.
Voltage and current results
The voltage and current results are defined as follows:
Result Description
RMS Square root of the mean of the square of the current
or voltage averaged over N cycles
Peak Highest measured magnitude value of the voltage or
current
Crest factor Peak value / RMS value
Frequency Frequency of the signal
Power results
The power in a system is described by several physical quantities: real power, reactive
power
, complex power, and phase angle. In Figure 19-1, you can see how these quan-
tities are related if the voltage and the current are sinusoidal signals.