Setting parameters:
<Frame>
<Field>
<Operator> ANY | OFF | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | RANGe
ANY = OFF
The index of the required pattern is not relevant.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
This condition requires one endpoint value to be set using
TRIGger<m>:
CMSB:ADVanced:IMIN.
INRange | RANGe
This condition requires a range of endpoint values to be set
using TRIGger<m>:CMSB:ADVanced:IMAX and
TRIGger<m>:CMSB:ADVanced:IMIN.
Usage: Setting only
24.17.7.3 Filter
BUS<m>:CMSB:FIL
Ter:ENABle <Enable>
Enables filtering for the custom serial bus.
Suffix:
<m>
.
1..4
Parameters:
<Enable> ON | OFF
*RST: OFF
BUS<m>:CMSB:FILTer:ERENable <ErrorName>, <Enabler>
Enables filtering on error
.
Suffix:
<m>
.
1..4
Setting parameters:
<ErrorName>
<Enabler> ON |
OFF
Usage: Setting only
BUS<m>:CMSB:FILTer:ERRor<n>:ENABle <Enable>
Enables filtering on error. Y
ou can filter all error types in parallel.