Instrument setup
R&S
®
RTO6
91User Manual 1801.6687.02 ─ 05
Enable
If enabled, each new data point in the diagram area remains on the screen for the
duration that is defined using Time, or as long as
Infinite persistence is selected.
If disabled, the waveform points are displayed only for the current acquisition.
Remote command:
DISPlay:PERSistence[:STATe] on page 1319
Infinite persistence
If infinite persistence is enabled, each new waveform point remains on the screen until
this option is disabled. Use infinite persistence to display rare events in the signal.
Remote command:
DISPlay:PERSistence:
INFinite on page 1320
Time
Sets a time factor that controls how long the waveforms points fade away from the dis-
play. Thus, the R&S RTO6 emulates the persistence of analog phosphor screens.
Remote command:
DISPlay:PERSistence:TIME on page 1320
Reset
Resets the display, removing persistent all waveform points.
Remote command:
DISPlay:PERSistence:RESet on page 1320
4.4.2 Signal settings
Access: "Menu" > "Settings" > "Display" > "Signal"
Display settings