1159User Manual 1801.6687.02 ─ 05
Arithmetic Reset > Mode
Defines when the envelope and average evaluation restarts.
"None"
No restart, the number of acquisitions considered by the waveform
arithmetics is not limited.
"Time"
Restarts the envelope and average calculation after the time defined
in "Time".
"W
aveforms"
Restarts the envelope and average calculation after a number of
acquired waveforms defined in "Count".
Remote command:
ACQuire:ARESet:MODE
on page 1335
ACQuire:ARESet:TIME on page 1335
ACQuire:ARESet:COUNt on page 1335
Reset
Forces the immediate restart of the envelope and average calculation for all wave-
forms.
Remote command:
ACQuire:ARESet:IMMediate on page 1335
Optimize scaling
Selects the scaling for the display of the results.
"Autoscale the signals"
Automatically selects the most appropriate scale for the display of the
results.
"Keep present settings"
The present display settings are not changed.
Remote command:
POWer:DONRes:AUTO on page
2535
Report comment
In this field you can write a comment that is displayed in the measurement report, for
example, a measurement description.
Execute
Starts the "Dynamic on Resistance" measurement.
Remote command:
POWer:DONRes:EXECute on page 2536
19.4.3 Slew rate
The "Slew Rate" analysis measures the rate of change of the voltage or current wave-
form during the switching of the switching transistor.
Required probes:
Differential voltage probe
Control loop