Remote control commands
R&S
®
RTO6
1833User Manual 1801.6687.02 ─ 05
Parameters:
<SignalName> String with the signal name as defined in the DBC file.
TRIGger<m>:CAN:SYMBolic:DMAX <DataPatternT
o>
Defines the maximum data value of the signal.
This value is required to specify a range if condition INRange
| OORange is set with
TRIGger<m>:CAN:DCONdition.
Parameters:
<DataPatternTo> Decimal representation of the data pattern
Range: -100E+24 to 100E+24
Increment: 1
*RST: 1
TRIGger<m>:CAN:SYMBolic:DMIN <DataPattern>
Defines the minimum data value of the signal.
To set the condition, use
TRIGger<m>:CAN:DCONdition.
Parameters:
<DataPattern> Decimal representation of the data pattern
Range: -100E+24 to 100E+24
Increment: 1
*RST: 0
TRIGger<m>:CAN:SYMBolic:SGEValue <SignalEnumValue>
Sets a symbolic data value for signals with enumerated values.
Parameters:
<SignalEnumV
alue> Numeric value according to the value definition in the DBC file
Example: Definition line in DBC file:
VAL_ 2175091489 Gear 0 "Idle" 1 "Gear_1" 2 "Gear_2" 3 "Gear_3"
4 "Gear_4" 5 "Gear_5" ;
Search for "Gear_4"
TRIGger:CAN:SYMBolic:SGEValue 4
Symbolic decode results
BUS<m>:CAN:FRAMe<n>:SDA
Ta?
Returns the complete symbolic data of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.