Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1123User Manual 1801.6687.02 ─ 05
The Setup settings, can be used by all measurements, like deskewing.
19.1.1.1 Auto deskew
The "Auto Deskew" dialog box guides you through the auto deskew of your current and
voltage probes.
Required equipment:
R&S
RT-ZF20 power deskew fixture
Rohde & Schwarz voltage probe
Rohde & Schwarz current probe
1. Select "Analysis" > "Power".
2. Under "General", select "Auto Deskew".
3. Connect the voltage probe and the current probe to the oscilloscope.
4. Connect the probes to the R&S RT-ZF20 power deskew fixture as shown in the
"Channels" tab: