Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1148User Manual 1801.6687.02 ─ 05
4. Deskew the probes as described in Chapter 19.1.1.1, "Auto deskew",
on page 1
123
.
5. Connect the probes to the DUT as shown in the "Channels" tab:
6. Select the correct channels for the "Current Source" and the "Voltage Source".
7. Select the "Settings" tab.
8.
Set the "Standard in use"
9. Set the "Frequency" according to your signal.
10. Select an"Optimize Scaling" option.
11. Tap "Execute".
On the screen you can see the measurement of the current, the voltage and the
power. Also there is a table giving information about important measurement
parameters. For details, see Chapter 19.3.3.1, "Current harmonic results",
on page 1146.
19.3.3.3 Current harmonic settings
In the "Channels" tab, you set the current source and the voltage source, see also:
Chapter 19.2.1, "Channels tab", on page 1134.
Input