Data bits
Sets the number of data bits of a word in a range from 5 bits to 8 bits. If no parity bit is
used, then 9 data bits are possible.
Remote command:
BUS<m>:UART:SSIZe on page 1731
Stop bits
Sets the number of stop bits: 1 or 1.5 or 2 stop bits are possible.
Remote command:
BUS<m>:UART:SBIT on page 1731
Parity
Defines the optional parity bit that is used for error detection.
"None"
No parity bit is used.
"Odd"
The parity bit is set to "1" if the number of data bits set to "1" is even.
"Even"
The parity bit is set to "1" if the number of data bits set to "1" is odd.
"Mark"
The parity bit is always a logic 1.
"Space"
The parity bit is always a logic 0.
"Don't care"
The parity is ignored.
Remote command:
BUS<m>:UART:PARity
on page 1730
13.4.2.2 Advanced settings
Access: [Bus] > "Advanced" tab.
Autoset
Starts software algorithms for determining the signal threshold levels and bitrate.