Remote control commands
R&S
®
RTO6
2557User Manual 1801.6687.02 ─ 05
POWer:TRANsient:TRGLevel <T
riggerLevel>
Parameters:
<T
riggerLevel> Range: -1000 to 1000
Increment: 1E-3
*RST: 0
Default unit: V
POWer:TRANsient:TRGSlope <TriggerSlope>
Sets the edge type for the trigger event.
Parameters:
<T
riggerSlope> POSitive | NEGative | EITHer
*RST: POSitive
24.23.16 Spectrum
POWer:SPECtrum:AUToscale....................................................................................... 2557
POWer:SPECtrum:EXECute......................................................................................... 2557
POWer:SPECtrum:FREQuency.....................................................................................2557
POWer:SPECtrum:REPort:ADD.................................................................................... 2558
POWer:SPECtrum:RCOunt?.........................................................................................2558
POWer:SPECtrum:RESult<m>:FREQuency?..................................................................2558
POWer:SPECtrum:RESult<m>:LEVel?...........................................................................2558
POWer:SPECtrum:AUToscale <Autoscale>
Selects the scaling for the display of the results.
Parameters:
<Autoscale> AUTO | MANUAL
*RST: MANUAL
POWer:SPECtrum:EXECute
Starts the spectrum measurement.
Usage: Event
POWer:SPECtrum:FREQuency <SmpsFrequency>
Sets the SMPS switching frequency.
Parameters:
<SmpsFrequency>
Range: 1 to 100E+6
Increment: 1
*RST: 1E+6
Default unit: Hz
Power analysis (option R&S RT
O6-K31)