Remote control commands
R&S
®
RTO6
2387User Manual 1801.6687.02 ─ 05
24.17.28 DDR (option R&S RT
O6-K91)
Configuration.......................................................................................................
2387
Decode results.................................................................................................... 2389
Search settings................................................................................................... 2393
Search results..................................................................................................... 2398
Eye diagram analysis..........................................................................................2402
24.17.28.1 Configuration
In all BUS<m>:DDR commands, the suffix <m> selects the serial bus.
BUS<m>:DDR:STANdard............................................................................................. 2387
BUS<m>:DDR:DATA:HYSTeresis.................................................................................. 2387
BUS<m>:DDR:DATA:SOURce...................................................................................... 2388
BUS<m>:DDR:DATA:THReshold...................................................................................2388
BUS<m>:DDR:STRBe:HYSTeresis................................................................................2388
BUS<m>:DDR:STRBe:SOURce.................................................................................... 2388
BUS<m>:DDR:STRBe:THReshold.................................................................................2389
BUS<m>:DDR:BITRate................................................................................................ 2389
BUS<m>:DDR:STANdard <Standard>
Selects the DDR standard.
Suffix:
<m>
.
1..4
Parameters:
<Standard> DDR3
DDR3: requires option R&S R
T
O6-K91.
*RST: DDR3
Usage: Asynchronous command
BUS<m>:DDR:DATA:HYSTeresis <DataHysteresis>
Sets a value for the hysteresis of the data.
Suffix:
<m>
.
1..4
Parameters:
<DataHysteresis> Range: 0 to 10
Increment: 0.01
*RST: 0.15
Default unit: V
Usage: Asynchronous command