Deembedding (option R&S RTO6-K121)
R&S
®
RT
O6
1200User Manual 1801.6687.02 ─ 05
3. Select the input channel to which the measurement setup is connected.
4.
Add the components of the measurement setup:
a)
Tap the "+" icon.
b) Select the component type.
The component is added to the setup, and it is disabled by default (bypassed). The
DUT and a probe or other input component are included by default.
5. Configure each component:
a) Tap the "Configure" button.
b) Enter the settings as required.
All settings are described in Chapter 20.2, "Deembedding settings",
on page 1200.
c) If necessary, tap "Open" and load the S-parameters from file.
As long as the configuration is incomplete or conflicting, the message "Invalid
configuration" is shown.
d) Enable the component to consider it during deembedding.
e) Tap the subtab of the next component, and configure it in the same way.
6. On the "Setup" tab, set the Effective bandwidth and the Delay compensation.
7. Enable the deembedding.
8. If you want to keep the configuration for later use, save it to file.
20.2 Deembedding settings
20.2.1 Measurement setup
In the "Deembedding" > "Setup" tab, you add all components of the measurement
setup and define all settings that are related to the complete path.
Deembedding settings